Standard set value
|
Current
|
Display
Limit |
Accuracy
|
|
Slow
|
Fast
|
|
100 mOhm ~ 999 mOhm
|
100mA
|
±99.9%
|
Within ±0.3% ± 1 digit
|
Within ±0.3% ± 2digit
|
|
1.00 Ohm ~ 9.99 Ohm
|
100mA
|
|
10.0 Ohm ~ 99.9 Ohm
|
10mA
|
|
100 Ohm ~ 999 Ohm
|
5mA
|
|
1.00 kOhm ~ 9.99 kOhm
|
500µA
|
|
10.0 kOhm ~ 99.9 kOhm
|
50µA
|
|
100 kOhm ~ 999 kOhm
|
5µA
|
|
1.00 MOhm ~ 9.99 MOhm
|
0.5µA
|
Within ±0.3% ± 1 digit
|
Within ±0.6% ± 2digit
|
|
10.0 MOhm ~ 100 MOhm
|
0.05µA
|
Within ±(setting
value/100+3%)± 2digit
|
----------------------------
|
|
(mOhm check)
|
100mA
|
0 ~ 999
mOhm |
Within ±0.2%
± 1 digit |
Within ±0.2% ± 2digit
|
| Measuring Way |
4 Terminals Measurement: below 99.9 Ohm, 2 Termianls
Measurement: 100Kohm over & contact check on or off |
| Judgement Setting Limit |
Both upper/low limit: ±0.00% ~ ±99.9%(mohm
check is 000mohm ~999mohm) |
| Input/Output Signal(photo-isolation) |
Input signal=START, HOLD; Output signal=HI/GO/LO(judged
result); EOC (Measurement finish) |
| Ambient Temperature |
0°C ~ 50°C |
| Ambient Humidity |
Below 85% |
| Outer Dimension |
250(W) x 85(H) x 345 (D) mm |
| Power Supply |
AC100V/117V/220V/240V; 50/Hz/60Hz; approx. 30VA |
| Weight |
4 kg approx. |
| Option |
Contact Check Function RS232 Printer output |