Duncan Instruments Canada Ltd
Duncan Instruments Canada Ltd.
Electronic Measurement Solutions since 1968
ISO 9001
 
1-877-742-4448
Products

 

UNI-SIM Vital Signs Simulator (NIBP / SPO2 and ECG)

Insulation and Continuity Tester
  • Accurate 6 in 1 vital signs simulation-NBP,SpO2,ECG,emp. respiration,IBP
  • Synchronised heart rate
  • Hand-held and Battery powered
  • On-board memory and keyboard
  • Manual and automatic operation
  • Onboard data entry and storage
  • Custom test protocols
  • Download / Upload software
Technical Specifications
Non-invasive Blood Pressure
Simulation Waveform

Oscillometric

Integrated Pump

0 to 350 mmHg user configurable

Leak Test

User configurable between 0-350mmHg

Chronometer

Configurable upto 999 secs

Digital Manometer

0 - 410 mmHg

Pressure Accuracy

+/- 0.5% FS

Oxygen Saturation

Range

50 to 100%

Accuracy

1 bpm

Chronometer

Test option to test response time SpO2 monitor

ECG Arrhythmia Simulator

ECG

Full 12 lead simulation including high level output

Wave Form

Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T

Rate 20 - 300 BPM typed value
Performance Waveforms Sine, Square, Triangle, and Pulse
Pacer Only Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial & Ventricular Pacer Heart rate 70 BPM, R Wave Detection
Arrhythmia Waveforms
Atrial Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Contraction
Atrial Conduction First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock
Ventricular Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Focal PVC
Respiration
Respiration Rates 5, 10, 15, 30, 60, 120, 180 Breaths per Minute
Base Resistances 250, 500, 750, 1000 ohms
Apnoea Simulation 0 60 seconds duration 0 300 seconds interval
Temperature
Simulation YSI 400 / 700 Static
Range Preset at 25, 33, 37and 41C
General Specification
Operation Battery cell, instiu charge
Battery Charger 110-230VAC, 50/60 Hz
Battery Life 4-24 hours testing depending on simulations
Memory Capacity Approximately 10,000 records
Communication Via Blue Tooth
Display Monochorme, 4/1 VGA full graphics
Keypad Alpha-numeric
Weight <1.5 kg
Size (L x W x D) 27- x 110 75 mm / 10.5 x 4 x 3 "
Operating Conditions 10-30C, 0-90% RH - NC
Storage Environment -15C - +60CC
Environmental Protection IP40
Accessories:
SIM Range Accessory Datasheet Download
Documentation:
Datasheet Download

 

Home | Products | Services | Software | Contact Us