IR Temperature |
Range |
-58 to 3992°F (50 to 2200°C) |
Accuracy |
-58 to 68°F (-50 to -20°C): ± 6.3°F(3.5°C) 68 to 932°F (20 to 500°C ): ±1% rdg. +1.8°F (1.0°C) 932 to 1832°F (500 to 1000°C): ±1.5% rdg. 1832 to 3992°F (1000 to 2200°C): ±2.0% rdg. |
Resolution |
0.1°F (0.1°C) |
Contact Type K T/C Temperature |
Range |
-58 to 2498°F (-50 to 1370°C) |
Accuracy |
-58 to 32°F (-50 to 0°C): ±4. 5°F (2.5°C) 32 to 2498°F (0 to 1370°C): ±0.5% rdg. + 2.7°F (1.5°C) |
Resolution |
0.1°F (0.1°C) |
Ambient Temperature |
Range |
32 to 122°F (0 to 50°C) |
Accuracy |
50 to 104°F (10 to 40°C): ±0.9°F (0.5°C) >104°F (40°C): ±1.8°F (1.0°C) |
Dewpoint and Wet Bulb Temperature |
Range |
32 to 122°F (0 to 50°C) |
Accuracy |
<40°C: ±0.9°F (0.5°C) >40°C: ± 1.8°F (1.0°C) |
Relative Humidity |
Range |
0 to 100% RH |
Accuracy |
0 to 20%: ±5% 20 to 40%: ±3.5% 40 to 60%: ±3% 60 to 80%: ±3.5% 80 to 100%: ±5% |
General Specifications |
Optical Resolution (D:S) |
50:1 |
Dual Laser |
Yes |
Spectral Response |
8 to 14µm |
Emissivity |
0.1 to 1.0 (Adjustable) |
Response Time |
150ms |
Display Size/Type |
2.2" TFT Color LCD |
Backlit Display |
Yes |
Display Hold |
Yes |
High/Low Alarms |
Yes (Audible (beep) and visual (on-screen)) |
Max/Min |
Yes |
Average and Differential |
Yes |
Datalogging Capabilities |
Yes |
F/C Switchable |
Yes |
Trigger Lock |
Yes |
Autoshut off |
Yes (adjustable 3,15,60mins) |
Internal Memory |
Yes (72mb), Expandable with micro SD card |
Thermocouple Input |
Yes |
Tripod Mountable |
Yes |
Laser Class |
Class II |
Low Battery Indicator |
Yes |
Power Supply |
Li-Ion rechargeable battery |
Battery Life |
Approx. 4 hours continuous use |
Battery Charge Time |
Approx. 2 hours |
Product Certifications |
CE, RoHS |
Operating Temperature |
32 to 122°F (0 to 50°C) |
Storage Temperature |
-4 to 140°F (-20 to 60°C) |
Operating Humidity Range |
10-90% |
Dimensions |
8.1 x 2.4 x 6.1" (205 x 62 x 155mm) |
Weight |
14.5 oz (410g) |